GEMMA · Membrane plasmapheresis
Configured for your protocol.
One or two needles, a total flow from 2 to 200 mL/min and the PFM-500 or PFM-800 filter your procedure calls for. Every parameter on this page carries its source.

- Device weight
- 8 kg
- Configurable total flow
- 2–200 mL/min
- Access modes
- One / two needles
- Dimensions
- 320 × 200 × 200 mm
GEMMA / The system
Membrane or centrifuge.
Two ways to separate plasma. GEMMA uses a membrane filter; the right-hand column describes centrifugal systems in general, which differ by model.
| Criterion | Membrane · GEMMA | Centrifuge | Source |
|---|---|---|---|
| Extracorporeal volume | Filter priming ≤ 20 mL (PFM-500) or ≤ 25 mL (PFM-800); single-needle draw 3–10 mL per cycle. Total circuit volume: confirm with our team. | Set by each model's separation chamber and lines. | MEMBRANA p. 10 · GEMMA p. 4 |
| Vascular access | One or two needles; peripheral venous access where clinical judgement allows. | Depends on the model and the protocol. | GEMMA pp. 3–4 |
| Procedure time | Depends on the protocol and the volume processed. | Depends on the protocol and the volume processed. | — |
| Space required | 320 × 200 × 200 mm, 8 kg. | Depends on the model; check its specification. | GEMMA p. 4 |
| Consumables | SM-PF-01 circuit and a PFM-500 or PFM-800 filter, single use. | A model-specific disposable set. | GEMMA pp. 4, 8 · MEMBRANA p. 1 |
| Mobility | Bedside, ICU, therapy room, ambulatory and mobile settings, as described by the manufacturer. | Depends on the model. | GEMMA pp. 2, 10–12 |
Centrifugal systems differ by manufacturer and model. Compare against the specification of the device you are evaluating.
GEMMA / Direct control
Your settings, on the equipment.
Flow, volumes and monitored pressure are visible directly on the equipment.

- FLOW
- Configurable total flow
- K
- Blood : anticoagulant ratio
- VOLUME
- Blood / anticoagulant volumes
- MODE
- Cycle status
- PRESSURE
- Monitored pressure
FLOW, K, VOLUME and MODE are indicator names in the brochure. PRESSURE identifies the monitored parameter here; it is not a newly claimed physical control label.
Manufacturer documentation · GEMMA · pages 4, 8, 10
GEMMA / Components
Every component, with its reference.
The filters and the circuit GEMMA uses, with their reference codes. Request a quote for the quantities your programme needs.

PFM-500
Membrane plasma filter, single use.
- Reference
- PFM-500
- Compatibility
- GEMMA with the SM-PF-01 / SM-PF-01U circuit
- Nominal filter flow
- 50 mL/min
- Maximum filter priming
- ≤ 20 mL
- Maximum filter inlet pressure
- 90 mmHg
Source: MEMBRANA p. 10

PFM-800
Membrane plasma filter, single use.
- Reference
- PFM-800
- Compatibility
- GEMMA with the SM-PF-01 / SM-PF-01U circuit
- Nominal filter flow
- 50 mL/min
- Maximum filter priming
- ≤ 25 mL
- Maximum filter inlet pressure
- 90 mmHg
Source: MEMBRANA p. 10

SM-PF-01
Sterile, single-use, non-pyrogenic blood circuit.
- Reference
- SM-PF-01
- Compatibility
- GEMMA · PFM-500 / PFM-800
- Access
- One / two needles
- Second stage
- SM-PF-01U, optional
Source: GEMMA pp. 3–4, 8
GEMMA / Documentation
Source documents.
The manufacturer documents behind every figure on this page, in their original language.
Before implementation
Practical questions.
Twelve questions clinical, technical and procurement teams ask before choosing GEMMA.
01Clinical integration
02Operation and training
03Consumables and continuity
Apheresis Technology / GEMMA
See GEMMA working in your clinic.
Book a demonstration. We review your clinical setting, technical requirements and the configuration that fits your protocols.
- 01
Technical brochure
- We ask for
- Your email
- You get
- The PDF, straight away.
- 02
Demonstration
- We ask for
- Four fields
- You get
- A call within 48 hours.
- 03
WhatsApp
- We ask for
- One click
- You get
- An immediate reply during business hours.