GEMMA · Membrane plasmapheresis

    Configured for your protocol.

    One or two needles, a total flow from 2 to 200 mL/min and the PFM-500 or PFM-800 filter your procedure calls for. Every parameter on this page carries its source.

    GEMMA with its blood circuit in a clinical setting
    GEMMA with its blood circuit in a clinical setting
    Device weight
    8 kg
    Configurable total flow
    2–200 mL/min
    Access modes
    One / two needles
    Dimensions
    320 × 200 × 200 mm

    GEMMA / Technical data

    Parameters, with their source.

    Specifications shown are based on manufacturer documentation. Configuration, electrical requirements, consumables and availability may vary by market. Confirm current specifications before procurement or commissioning.

    01Fluid control

    ParameterValueSource
    Configurable total flow2–200 mL/minGEMMA · p. 4
    Flow increment: 2–50 mL/min1 mL/minGEMMA · p. 4
    Flow increment: above 50–200 mL/min5 mL/minGEMMA · p. 4
    Blood : anticoagulant ratio5:1–25:1GEMMA · p. 4

    02Access & pressure

    ParameterValueSource
    Displayed pressureUp to 300 mmHgGEMMA · p. 4
    Single-needle return pressure60–240 mmHgGEMMA · p. 4
    Dual-needle pressure threshold60–300 mmHgGEMMA · p. 4
    Single-needle draw volume3–10 mLGEMMA · p. 4
    Access modesOne / two needlesGEMMA · p. 4

    03Physical & operating environment

    ParameterValueSource
    Device weight8 kgGEMMA · p. 4
    Dimensions320 × 200 × 200 mmGEMMA · p. 4
    Operating temperature10–35 °CGEMMA · p. 4
    Relative humidity≤ 80%GEMMA · p. 4

    04Electrical & continuity

    ParameterValueSource
    Typical apparent-power consumption≤ 60 VAGEMMA · p. 4
    Reference electrical input220 V / 50 HzGEMMA · p. 4
    Continuous operation> 8 hGEMMA · p. 4
    Specified MTBF5 000 hGEMMA · p. 4

    3–10 mL is the volume per single-needle draw / cycle. It is not the total extracorporeal volume. Filter priming and maximum filter inlet pressure are separate values in the component catalogue.

    Published operation and MTBF values are brochure specifications, not a guarantee of reliability or service availability.

    Manufacturer documentation · GEMMA · pages 4, 11–12

    GEMMA / The system

    Membrane or centrifuge.

    Two ways to separate plasma. GEMMA uses a membrane filter; the right-hand column describes centrifugal systems in general, which differ by model.

    Membrane and centrifuge: two separation principlesMembraneBloodCellular componentsPlasmaSeparated by pore sizeCentrifugePlasmaRotationCellular componentsSeparated by density
    Membrane and centrifuge: two separation principles Schematic principles, not to scale. GEMMA separates by membrane.
    CriterionMembrane · GEMMACentrifugeSource
    Extracorporeal volumeFilter priming ≤ 20 mL (PFM-500) or ≤ 25 mL (PFM-800); single-needle draw 3–10 mL per cycle. Total circuit volume: confirm with our team.Set by each model's separation chamber and lines.MEMBRANA p. 10 · GEMMA p. 4
    Vascular accessOne or two needles; peripheral venous access where clinical judgement allows.Depends on the model and the protocol.GEMMA pp. 3–4
    Procedure timeDepends on the protocol and the volume processed.Depends on the protocol and the volume processed.—
    Space required320 × 200 × 200 mm, 8 kg.Depends on the model; check its specification.GEMMA p. 4
    ConsumablesSM-PF-01 circuit and a PFM-500 or PFM-800 filter, single use.A model-specific disposable set.GEMMA pp. 4, 8 · MEMBRANA p. 1
    MobilityBedside, ICU, therapy room, ambulatory and mobile settings, as described by the manufacturer.Depends on the model.GEMMA pp. 2, 10–12

    Centrifugal systems differ by manufacturer and model. Compare against the specification of the device you are evaluating.

    GEMMA / Direct control

    Your settings, on the equipment.

    Flow, volumes and monitored pressure are visible directly on the equipment.

    Original interface detail from the GEMMA brochure.
    Original interface detail from the GEMMA brochure.
    FLOW
    Configurable total flow
    K
    Blood : anticoagulant ratio
    VOLUME
    Blood / anticoagulant volumes
    MODE
    Cycle status
    PRESSURE
    Monitored pressure

    FLOW, K, VOLUME and MODE are indicator names in the brochure. PRESSURE identifies the monitored parameter here; it is not a newly claimed physical control label.

    Manufacturer documentation · GEMMA · pages 4, 8, 10

    GEMMA / Components

    Every component, with its reference.

    The filters and the circuit GEMMA uses, with their reference codes. Request a quote for the quantities your programme needs.

    • PFM-500

      PFM-500

      Membrane plasma filter, single use.

      Reference
      PFM-500
      Compatibility
      GEMMA with the SM-PF-01 / SM-PF-01U circuit
      Nominal filter flow
      50 mL/min
      Maximum filter priming
      ≤ 20 mL
      Maximum filter inlet pressure
      90 mmHg

      Source: MEMBRANA p. 10

    • PFM-800

      PFM-800

      Membrane plasma filter, single use.

      Reference
      PFM-800
      Compatibility
      GEMMA with the SM-PF-01 / SM-PF-01U circuit
      Nominal filter flow
      50 mL/min
      Maximum filter priming
      ≤ 25 mL
      Maximum filter inlet pressure
      90 mmHg

      Source: MEMBRANA p. 10

    • SM-PF-01

      SM-PF-01

      Sterile, single-use, non-pyrogenic blood circuit.

      Reference
      SM-PF-01
      Compatibility
      GEMMA · PFM-500 / PFM-800
      Access
      One / two needles
      Second stage
      SM-PF-01U, optional

      Source: GEMMA pp. 3–4, 8

    Full component catalogue

    Before implementation

    Practical questions.

    Twelve questions clinical, technical and procurement teams ask before choosing GEMMA.

    01Clinical integration

    02Operation and training

    03Consumables and continuity

    Apheresis Technology / GEMMA

    See GEMMA working in your clinic.

    Book a demonstration. We review your clinical setting, technical requirements and the configuration that fits your protocols.

    1. 01

      Technical brochure

      We ask for
      Your email
      You get
      The PDF, straight away.
    2. 02

      Demonstration

      We ask for
      Four fields
      You get
      A call within 48 hours.
    3. 03

      WhatsApp

      We ask for
      One click
      You get
      An immediate reply during business hours.
      Message us on WhatsApp
    Technical support